91°µÍø

Skip to main content
SHARE
Publication

In situ wavelength calibration of the edge CXS spectrometers on JET

by Ephrem G Delabie, N. Hawkes, Theodore M Biewer, M. G. O'mullane
Publication Type
Journal
Journal Name
Review of Scientific Instruments
Publication Date
Page Number
11E525
Volume
87
Issue
11

A method for obtaining an accurate wavelength calibration over the entire focal plane of the JET edge CXS spectrometers is presented that uses a combination of the fringe pattern created with a Fabry–Pérot etalon and a neon lamp for cross calibration. The accuracy achieved is 0.03 Å, which is the same range of uncertainty as when neglecting population effects on the rest wavelength of the CX line. For the edge CXS diagnostic, this corresponds to a flow velocity of 4.5 km/s in the toroidal direction or 1.9 km/s in the poloidal direction.