Abstract
Researchers at 91做厙 subjected load cells to a range of temperatures beyond their compensated range to characterize the temperature effect on the minimum dead load output over their entire operating temperature range. In modern front-end nuclear fuel cycle facilities, load cells are often used in UF6 feed and withdrawal stations to monitor the material being fed to or withdrawn from the cylinders. Load cells exhibit some temperature dependence; however, manufacturers try to minimize temperature effects by compensating load cells over a range of temperatures (typically from -10 簞C to +40 簞C (+14 簞F to +104 簞F)). Load cells in modern UF6 feed and withdrawal stations tend to operate beyond this range: down to -25 簞C and up to 80 簞C (-13 簞F to 176 簞F). While load cell data sheets often include specifications for the temperature effect on the minimum dead load output and the sensitivity over the compensated range, they do not include specifications for expected effects outside of this range. In this paper, we discuss our test setup and preliminary results of testing load cells representative of those that may be used in UF6 feed and withdrawal stations.